Supplied with Two Filters (Na and K), Air Supply Unit, Atomizer Tube, Sample
Beaker (06 Nos.), Gas Lighter, LPG Tube, Power Cord, Operational &
Instruction Manual with SOP, Calibration Test Certificate
(Without Traceability Document) and Warranty Card.
Technical Specification:
Element Range : Na, K, & Li : 0.1-100 ppm, Ca : 15-100 ppm, Ba : 50-1000 ppm
Sensitivity : Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm, Ba: 50 ppm
Filter Selection : Automatic by soft touch key
Resolution : 0.1 ppm
Detector : Silicon Photodiode
Unit of Measurement : ppm
Reproducibility : <1% Coefficient of Variation (CV) for 20 consecutive samples using 10 ppm Na set as
maximum standard “Curve Fitting Accuracy <1% error when 3 ppm Na/K and 5 ppm Li
are set as maximum standards.
Quick Calibration Facility : With Highest standard
Method : 40 method with View, Print & Delete facility
Calibration : Up to 05 standards per Elements with Curve Fitting Software in-built with
Instrument. (Automatic calculation of curve coefficient)
Measurement of Sample : Multi elements in single aspiration OR one by one.
Ignition System : Automatic
Flame Type : LPG & Dry Oil Free Air
Auto Flame ON / OFF Detection : Yes Audio and Visual Alarm
Display : 20 x 4 Line Alphanumeric LCD with Backlit
Keypad : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
Gas Control : Adjustable with Regulator
Atomizer : Axial Flow Type
Air Supply Unit (Separate Unit) : Consist of Air Compressor, Pressure Gauge, Pressure Regulator, Moisture Filter,
PU Tube (2 Meters) and Air filter with moisture absorbent catalyst
Error Messages : “LPG On/Off”, “Wrong Standard”, “Home Detection”, “Memory Full”, “Flame
Not ON”, “Value Out of Limit”, “Do Warm-up” & “In Correct Standard”.
Calibration Data Storage : Non Volatile Memory which can store, print & view report date & time, company name,
calibration date & time, calibration done by, standards results, calibration one on,
instrument serial number, instrument ID etc.
Searching of data : with record number, batch number, Date and by scrolling
Report Formation : As per GLP Compliance
Analysis Data Storage : More than 750 analysis data can be stored with corresponding calibration report of
lower & higher standard, company name, analysis date & time, record number, batch /
ID number, sample name, user name, result, dilution factor, analysis done on,
instrument serial number, instrument ID, method number, etc. Secondly, analysis data
can print directly from analysis mode without saving analysis data in instrument.
Password Protection : Date & time, calibration mode & delete stored analysis data
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
Printer : 40 / 80 Column Dot Matrix Printer (Printer have to procure optionally)
Power Supply : 230 VAC ± 10%, 50 Hz