Supplied with Combination Glass pH Electrode (0-70ºC), Standard Buffer Tablets (7.00 pH, 4.01 pH & 9.18 pH), Temperature Probe, Electrode Stand consists of Clamp, Base Plate & Rod, Dust Cover, DC Adaptor, ANALABsafe software, USB Interface Cable, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.
Technical Specification:
pH
Range : -2.000 to 20.000pH
Selectable Resolution : 0.1 / 0.01 / 0.001 pH
Relative Accuracy : ± 0.002 pH
mV/ORP
Range – mV / RmV : ± 2000.0
ORP : EH auto calibration to standard hydrogen electrode
Resolution : 0.1
Relative Accuracy : ± 0.2 mV OR 0.05%, whichever is greater.
Temperature
Range : -5.0 to 130.0ºC
Resolution : 0.1ºC
Relative Accuracy : ± 0.1ºC
Temperature Compensation : Auto / Manual
Temperature Calibration : On site by users with in-built software
Selectable Slope : 1) 80-120 %
2) Customized slope (User Selectable)
Total Numbers of Slope : 04 slope for Five-point calibration & 02 slope for Three-point calibration
Auto Buffer Recognition : 1) 4.01, 6.86 / 7.00, 9.18 / 10.01 pH @ 25ºC for Three points
2) 1.68, 4.01, 6.86 / 7.00, 9.18 / 10.01 & 12.45 pH @ 25ºC for five points
Buffer Standard : US / NIST / EURO / DIN
Error Messages : “Bad Buffer / Electrode”, “Temperature Probe”, “In-complete Calibration”,
“Calibration Due”, “Low Memory” & “pH Over Range”.
Display : 20 x 4 Line Alphanumeric Backlit LCD
Keypad : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
Display Reading : Simultaneous - pH, mV & Temperature
Short Link : Provided to diagnose the problem either with Instrument OR Electrode.
Calibration Reminder Alarm : 0-99 hours
Result hold / un-hold : By “hold / un-hold” Key
Calibration Data Storage : Non Volatile Memory which can store, print & view calibration date & time, calibration
done by, calibration point, pH, mV, temperature, % slope, validation pH for 3/5 points
five standards buffers, instruments serial number, instruments ID, calibration done on
etc.
Searching of data : with record number, batch number, Date and by scrolling
Report Formation : As per GLP Compliance
Analysis Data Storage : More than 1500 analysis data, with corresponding calibration data with company
name, analysis date & time, record number, batch number / ID, sample name, user
name, pH, mV, temperature reading, instrument serial number, instrument ID, analysis
done on etc. Secondly, analysis data can print directly from analysis mode without
saving analysis data in instrument.
21 CFR Part-11 Compliance : ● In-built with system
• 1 Admin and 7 users with different Login ID & Password.
● Audit trail can store more than 5000 lines in system itself.
● Audit trail includes action taken, date & time, action taken by whom.
● Audit trail can be transfer to PC through ANALABsafe software and
can print on 40 / 80 column dot matrix printer for record.
● Once Audit trail transfer to the PC or Printer, automatically system
will delete Audit trail records and before final deleting, the system
will give warning twice.
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
PC Interface : USB Type - B
Printer : 40 / 80 Column Dot Matrix Printer (Printer have to procure optionally)
PC Connectivity : By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in
GLP format, Calibration data and Audit trail from Instrument to PC in non-editable
format and is password protected.
Power Supply : DC Adaptor