Supplied with 250ml Reaction Vessel, Dual Platinum Combined Electrode, Teflon Paddle, Teflon Assembly, Moisture Trap (02 Nos.), 250ml Amber Reservoir Bottle, Dispensing Tube (02 Nos.), Micro tip (02 Nos.) Teflon Tube for Dispensing Path, Power Cord, ANALABsafe software, USB Interface Cable, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.
Technical Specification:
Method of Detection : Conductometric / Volumetric
Range of Moisture Detection : Better than 50 ppm to 100% (depends upon type of sample)
Display of Reading : Moisture in ppm, percentage (%), mg/ml of water, gm/L, Mol/L
Display : 20 x 4 Line Alphanumeric Backlit LCD
Magnetic Stirrer : Built-in with soft touch Key speed controller
Keypad : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
Method : Three pre-assigned method
1) liquid by volume,
2) liquid by weight
3) for solid sample
Method No. 4-40 are customized with View, Print & Delete facility.
Calibration : Average factor can be calculated by
1) water by volume,
2) water by weight
3) By DST
Stirring Time : 00:00 to 99:00 (mm : ss) – users adjustable
End Point Time : 1-20 seconds - users adjustable
Auto Neutralization : On / Off - users selectable - reduces the time between analyses & gives accurate results
Material of Construction : Epoxy Based Powder Coated Metallic Body
Type of Electrode : Dual Platinum - Combined Type
Type of Dispenser : Motorized - For accurate measurement
OVER TITER Facility : Audio / Visual Alarm at 15 ml to save KF Reagent
MOC of Dispenser : SS-316L & Teflon for Corrosion Protection & Maintenance Free Operation
Dispenser Calibration/Validation : Onsite by users
Flushing Mode : For cleaning dispensing path
Teflon Tube : To transfer KF reagent from KF reagent bottle to dispenser & dispenser to
reaction vessel
Auto Drain Facility : For cleaning reaction vessel - useful for large & multiple volume sample
Input : Electrode with Banana Connector
Short Link : Provided to diagnose the problem either with Instrument OR Electrode
End Point Detection : By Audio & Visual Alarm
Calibration Reminder Alarm : 0-99 Hours
Calculation of % RSD : Possible
Average KF Factor Calculation: Automatic – Maximum 10 factor can be stored in memory
Weight Entry : In Resolution of 0.1 mg / 0.01 mg – users Selectable
Weight Transfer : Possible through RS-232 Port (From Balance to AKF)
Balance Interface : Yes with RS-232 Port
Calibration Data Storage : Non-Volatile Memory which can store, print & view calibration date & time, calibration
method, calibration done by, stirring time, average KF factor, calibration done on,
instrument serial number, instrument ID etc.
Searching of data : with record number, batch number, Date and by scrolling
Report Formation : As per GLP Compliance
Analysis Data Storage : More than 1500 analysis data, with corresponding calibration data company, name,
analysis date & time, record number, batch / ID number, sample name, user name,
sample type and method, sample volume (ml / mg), sample density (for liquid sample),
stirring time, end point time, end point in ml, moisture result in %, ppm & H2O (mg /
ml), analysis done on, instrument serial number, instrument ID etc. Secondly, analysis
data can print directly from analysis mode without saving analysis data in instrument.
saving analysis data in instrument.
21 CFR Part-11 Compliance : ● In-built with system
• 1 Admin and 7 users with different Login ID & Password.
● Audit trail can store more than 5000 lines in system itself.
● Audit trail includes action taken, date & time, action taken by whom.
● Audit trail can be transfer to PC through ANALABsafe software and
can print on 40 / 80 column dot matrix printer for record.
● Once Audit trail transfer to the PC or Printer, automatically system
will delete Audit trail records and before final deleting, the system
will give warning twice.
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
Printer : 40 / 80 Column Dot Matrix Printer (Printer have to procure optionally)
PC Connectivity : By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in
GLP formate, Calibration data and Audit trail from Instrument to PC in non editable
formate and is password protected.
Power Supply : 230 VAC ±10%, 50 Hz