Automatic Programmable Melting Point Apparatus (21 CFR Part-11 Compliance)

Supplied with Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube, Sample Tapping Rod, Two Melting Standards 49ºC and 238ºC, Forceps, Power Cord, ANALABsafe software, USB Interface Cable, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.

Read more

Technical Specification:

Temperature Controller                : Microcontroller Based

Detection Principal of Melting Point : Automatic by Silicon Photodiode

Heating Media                            : By Aluminum Block

Temperature Range                     : +5°C above ambient to 400°C

Readability                                 : 0.1°C

Accuracy of Temperature              : A) 10°C above ambient to 100°C………… ± 0.2°C

  B) 101°C to 200°C ………… ± 0.4°C

  C) 200°C to 400°C …………… ± 0.5°C

Overheating Protection             : In-built - Set by user in method parameter

Heating Before EMP*                   : Automatic heating depending upon temperature difference

Heating Rates                             : 0.1 to 20.0ºC/min (Increment of 0.1ºC) - Adjustable

Detection of Melting                     : A) Melting Point B) Melting Range

  C) % Level of Melting Point between Melting Range

Type of Detection                        : Auto / Manual

Melting Point Average Reading : Display of average reading for three same sample

Melting Capillary                      : Three

Melting Channel                       : Three different sample analysis of different temperature in single run

Glass Melting Capillary                  : OD : 1.4 to 1.6mm | ID : 1.0mm | Length : 100mm | One End Sealed

Sample Filling Height                    : 3mm

Detection of Boiling Point         : Manually by viewing in TFT Display

Boiling Tube                             : One

Method                                      : 40 methods for Melting & 10 for Boiling sample with View, Print & Delete facility

Calibration                                 : Seven Calibration Standards (As per Pharmacopeia AND / OR customer's known

  standards.)

Maximum Heating Time                : Around 6 minutes from 50°C to 350°C

Maximum Cooling Time                : Around 6 minutes from 350°C to 50°C

Temperature Sensor                    : Duplex PT-100, One for internal used and One for calibration purpose by external

  source

Magnification                              : 5X

Keypad                                      : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS

Display                                      : 20 x 4 Line Alphanumeric Backlit LCD

Recoding of Melting Capillary   : Camera with TFT Display for viewing and Recoding of Melting Capillary in

  CAPillaryVIEW (Memory 64 GB)

Calibration Data Storage               : Non-Volatile Memory which can store, print & view company name, standard name,

  standard MP, observed MP, Calibration date and time, last calibration date & time,

  calibration done by, calibration status, instrument ID, serial number, model number etc.

Searching of data                        : with record number, batch number, Date and by scrolling

Report Formation                        : As per GLP Compliance

Analysis Data Storage                   : More than 1500 analysis data can be stored with analysis date & time, record

  number, Batch number / ID, sample name, user name, assigned method number, MP1#

   & MP2# reading of all three reading, lower and higher calibration standard data,

  instrument ID, serial number, model number, make, etc. Secondly, analysis data can

  print directly from analysis mode without saving analysis data in instrument.

21 CFR Part-11 Compliance       : ● In-built with system

  ● 1 Admin and 7 users with different Login ID & Password.

  ● Audit trail can store more than 5000 lines in system itself.

  ● Audit trail includes action taken, date & time, action taken by whom.

  ● Audit trail can be transfer to PC through ANALABsafe software and  

     can print on 40 / 80 column dot matrix printer for record.

  ● Once Audit trail transfer to the PC or Printer, automatically system

     will delete Audit trail records and before final deleting, the system

     will give warning twice.

Printer Interface                          : Centronics Parallel Port / Serial (RS-232 C) Port

PC Interface                               : USB Type – B

Printer                                       : 40 / 80 Column Dot Matrix Printer (Printer have to procure optionally)

 

 

PC Connectivity                           : By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in

  GLP format, Calibration data and Audit trail from Instrument to PC in non editable

  format and is password protected.

Power Supply                              : 230 VAC ±10%, 50 Hz, 150W

ANALAB SCIENTIFIC INSTRUMENTS PVT. LTD

ANALAB SCIENTIFIC INSTRUMENTS PVT. LTD
Manufacturer
India

Primary focus area

Analytical Products & Services

Add to Favourite
Book a Demo