Chrom Plus GC-4000 series

  • Versatile
  • Precise
  • Speed
  • Multiple Detectors(Max 4nos)  to cover wide range of chromatography applications 
  • Delivers high analytical performance
  • Electronic pneumatic controls for precise analysis
  • Computer controlled 
  • Dedicated capillary injection ports 

Read more

COLUMN OVEN

  • Size : H 290 mm x W 280mm x D 173mm
  • Volume : 14 Ltrs
  • Temperature : 5°C above the ambient to 450°C
  • Ramp Rate : 1°C to 70 °C per minute
  • No. of Ramps : 10 with storage of 20 methods
  • Cooling Rate - 250°C to 5 0°C within approx. 4 minutes at 25°C room
  • Temperature Gradient : 2°C max (on 200 mm dia circumference, 40mm from the sensor)
  • Temperature Accuracy : + 1 % of set value from 50° to 450°c 5 Independent Heated Zones 450°C

PNEUMATICS

  • EPPC ( electronic programmable pressure controller) for Carrier gas, Hydrogen, Air and Make-up gas.
  • Pressure switch provided to cut heating of oven, injector, detectors heater power in the event of carrier gas flow depletion
  • Differential Flow Controllers for carrier gas for Packed Column
  • Constant flow or Constant Pressure facility for split/splitless injector port .upto 0 to 1000 kPa (0 to145 psi )
  • Psi/kPa/bar pressure units conversion facility for supported carrier gas N2,H2,He,Ar
  • Split ratio : 9999:1

INJECTION SYSTEM

  • Single/dual Split/split-less (capillary column) injector port or single/ dual packed column injection port
  • Temperature range settable up to 450°C. in the steps of 1°C

DETECTOR 

FID 

  • High sensitivity flame ionization detector with auto flame ignition, Temperature range settable up to 450°C. in the step of 1 °C
  • Dynamic range : 107
  • Minimum Detection Level : <1.2pg C/sec of dodecane

TCD 

  • Dual path four Tungsten-Rhenium filaments
  • Temperature range settable up to 400°C in the step of 1 °C
  • Filament current range : 0-400 mA
  • Minimum Detection Level : 10 uv/ppm(vol) toluene
  • Filament temperature 150 - 450 °C in steps of 75°C with
  • Filament protection circuit

NPD/FTD/TID 

  • Temperature range settable up to 450°C. in the step of 1 °C
  • Minimum Detection Level : <0.1pg N/S C/sec (azobenzene) selective for nitrogen &<0.01pg P/S C/sec (malathion) selective for phosphorus.