Automatic particle counting is accepted throughout the world as an easy and accurate tool to detect contamination in solvents, aqueous solutions & other liquids.
Pamas GmbH of Germany is designing and manufacturing “Automatic Particle Counters” specially for applications in
Pharma and Chemical industry.
Even though all Pamas Particle Counters can be controlled by Pamas software they also may be used as “Stand Alone” instruments without external computer control.
Pamas supplies special software for data presentation and storage according to all commonly used standards.
Particle Counting System for Infusion Solutions and Clean Fluids.
The PAMAS SVSS is a standard particle measuring system for the measurement of clean, low viscous fl uids and of pharmaceutical liquids. The laboratory instrument is designed especially for the particle analysis of low viscous, aqueous solutions.
Applications:
• Infusion solutions
• Clean fluids
• Water analysis
• Water for Injection (Offshore)
• Suspensions
• Partscleaning application
• Size Distribution Analysis
Display shows measuring
results in real time
• Measurement of small volumes
down to 1 ml ampoules
• Large dynamic size range
• Simpliest operation
• Highest particle concentrations up to 200,000 p/ml
Pamas Small Volume Syringe System
USP, BP, EP, JP, KP, IP & many more
Sampling
The user can process multiple samples and observe the particle concentration in up to 16 (optionally32) free selectable size channels. Calibrating with polystyrene latex spheres of different sizes, the user can establish a calibration curve over the full range. An operator selectable flushing volume allows the user
Product features
• Preadjustable and automated sampling
• 16 free adjustable channels for particle size determination (32 channels on request)
• Size of sample container 1 ml - 2,000 ml.
• Analysis volume:100 μl - 1,000 ml
• Constant sample flow and exact volume control via stepper-motor-driven syringe
• Interchangeable syringes with 1 ml, 10 ml and 25 ml volume
SVSS - SMALL VOLUME KIT
There is an optional „Small Volume Kit“ available, consisting of:
Easy Contamination control of IV liquid in plastic bags.
Measuring a sample straight out of it’s standard packaging will give best results as an secondary contamination can hardly occur.There is no need of cleaning of beakers and storing them in dust free cupboards. Even if you have the most intricately task’s Pamas engineers will help find a solution.
SVSS - Special uses
Even very strongly colored liquids may be measured by using some simple tricks.
Important is that the color is not made by color pigments, which means that the color is made up of submicron colored particles.
Pamas engineers will solve this task by
1.choosing the Sensor model with smallest sensor cell depth.
2.Changing the sensor to a different laser voltage range.
3.Recalibrating the sensor with the new parameters.
4.Commencing with measuring.