Standard Features: Panel Construction Tongue and groove type PUF insulated sandwich panel with cam lock. The Equipment interior is constructed with stainless steel 304 grade finish-4 and exterior is constructed with pre coated GI sheet/ stainless steel 304 grade finish-4. Tray and Rack Duly electro polished stainless steel wire mesh tray and heavy duty square pipe rack with adjustable tray location. Door Heavy duty toughened double glass door installed with imported hinges and lock for easy operation. Backup System Equipment is installed with PLC based auto change over backup system for Refrigeration system and Steam Generator system. PLC based control system Advanced PLC program with built-in Auto-Tuning, PLC Processor based computing and Interactive Easy-to-Use HMI. Documentation • Equipment : DQ, IQ, OQ with PLC Validation, PQ-Conforming to ISPE recommendations, User Manual • Software : IQ, OQ, User Manual Capacity started from 400Ltrs. to 50000Ltrs. Measured with high accuracy data-logger at sample storage area with standard Set Point. Custom built Chambers are available in any dimensions and capacities.
Standard equipment's operating conditions 25°C / 60%RH 30°C / 65%RH 40°C / 75%RH 30°C / 75%RH Standard equipment's operating conditions 25°C / 40%RH 30°C / 35%RH 40°C / 25%RH Temperature Range : 20°C to 60°C Accuracy ≤ +/- 0.2°C Uniformity ≤ +/- 1.0°C Rel. Humidity Range : 40%RH to 90%RH Accuracy ≤ +/- 1.0% Uniformity ≤ +/- 3.0%